- Direct exposure to the air and room temperature are detrimental to human embryos and egg cells.
Indeed, various studies indicate that maintaining a survival environment is essential focontrol range
: by time, days, & weeksr the success rate of pregnancy.
Organs of FS-1 controls the suitable temperature, CO2 level for the screening egg cells or embryos,
and other procedures.
- Easily movable, roomy enough to fit any kinds of microscope including the stereo microscope.
- Effective magnetic revolving open/close system for each opening to prevent the air flow from outside.
- Lockup holder to fix a CO2 tank in a position.
- Polycarbonate or reinforced acrylic plate arecustomizable for the front transparent panel.
(Standard type: polycarbonate plate)
- Work surface is made of stainless steel for easy clean with alcohol or H2O2.
- "Automatic Timer" to set time, date, and duration. (days, weeks, months, and years)
- "IVF Environmental Control System," the touch screen panel to control the air temperature,
humidity and CO2 levels. (Separate on/off switches for CO2, humidity, and HEPA filter)
- "TouchBright" for the LED control.
- HEPA filter to eliminate any minute particles of dust in the chamber.
- Able to control and record the temperature, gas concentration/ gradient over time-
with Chamlide CCP-IVF software or MetaMorph software.
- Cleaning the touch screen panel with alcohol to terminate any contamination.
- Touch screen monitor to display microscopic camera images
- Separable humidifying module inside of the chamber to prevent contamination
- Able to fill up the humidifier from outside
- Silicon slits to connect to the microscope or other electric devices from outside
- Height changeable of the unit for the operator in either sit-up or stand-up position
depending on the request.
Specification
Exterior size
1,200mm (W) X 560mm (D) X 1,616mm (H)
(Any size and shapes are available)
Working interior size
1,150mm (W) X 535mm (D) X 651.5mm (H)
Temperature control
- Temperature control method: PID
- Heating method: Air temperature control using a cartridge heater
- Temperature range: ambient +3℃ ~ +45℃
- Step: 0.1 ℃
- Sensor type: PT 100 ohm
- Blower fan power consumption: 20W
- Cartridge heaters power consumption: 2kW
- Time to equilibrate inside: <1hr
External sensor
- Temperature range: ambient 0℃ ~ 100℃
- Sensor type: PT 100 ohm type
- Length: 1m
Co2 Control
- Control range: 1% ~ 15%
- Sensor type: NDIR CO2 sensor
- PID solenoid valve control
- Housing gas or tank are available
- Time to reach at set value: < 10 min
- Time to equilibrate inside: <10 min
Humidity control
- Control range: ~ 99% .
- Sensor type: integrated circuit humidity sensor
- Time to reach at set value: < 0.5 hr
- Time to equilibrate inside: <1hr
- Spray method: ultrasound evaporation
- Water bottle capacity: 4.8L
Touch screen control panel Humidifier and Water bottle
Automatic electric power control
Control rage: by time, day, and week
Touch screen control panel
Touch screen monitor for camera display or software
Hight adjustable option for sitting or standing O2 control option
- Control range: 1% ~ 100%
- Tube capacity : 17Ø x ≥ 100mm test tube
- Heating method : Thin layer heater
- Heating method : Thin layer heater
- Sensor : PT 100 ohm
- Temperature control : PID method
- Material : Silver anodized aluminum alloy
Incubator material
- Acrylic plastic, polycarbonate plastic, and steel
Optional wide table for specimen dishes
Model No. Description
FS-R-10 single IVF environmental chamber for 1 operator
FS-R-12 double IVF environmental chambers for 2 operators
(in the front and the back)
- Customized IVF chamber of any other functions, sizes or shapes are available upon request.